Home/Standards/NFPA 99 (2024)

Standard

NFPA 99 (2024)

The Health Care Facilities Code. It used to require conductive flooring in operating rooms. That requirement came out in 2015.

2.5 x 10^4 to 1.0 x 10^6ohmsNFPA 99 (historical, pre-2015 editions)

What it means for your floor

01

What changed and when

NFPA 99 historically required conductive flooring in the 2.5 x 10^4 to 1.0 x 10^6 ohm range in operating rooms using flammable anesthetics. That requirement was removed in the 2015 edition and is still absent from the current 2024 edition.

02

Why it was removed

Flammable inhaled anesthetics are essentially gone from US surgical practice, so the ignition risk the requirement existed to control went with them.

03

What this means if somebody cites it to you

NFPA 99 still governs health care facility construction broadly. It simply no longer carries this resistance range. If a specification cites NFPA 99 as the reason for a conductive floor, it is worth going back to the source before you buy to that band.

Sources

    NFPA 99 (2024)what it asks of a floor